Step 8
Press the scroll key until the display shows . Send a 10V
signal to terminals 15 and 16 with the correct polarity. Press the
scroll key for at least 3 seconds. The display will blink for a
moment until a new value is obtained. If the display didn’t blink
or if the obtained value is equal to -199.9 or 199.9, then calibra-
tion failed.
Perform step 9 to calibrate mA function ( if required ) for input
2.
Step 9
Press the scroll key until the display shows . Send a
20mA signal to terminals 15 and 16 with the correct polarity.
Press the scroll key for at least 3 seconds. The display will blink
for a moment until a new value is obtained. If the display didn’t
blink or if the obtained value is equal to -199.9 or 199.9, then
calibration failed.
• Perform step 10 to calibrate offset of cold junction compensa-
tion, if required. Set the DIP switch for T/C input.
Step 10
Set up the equipment according to
the following diagram to calibrate
the cold junction compensation.
Note that a K type thermocouple
must be used.
The 5520A calibrator is configured for K type thermocouple out-
put with internal compensation. Send a 0.00°C signal to the unit
under calibration. The unit under calibration is powered in a still-
air room with a temperature of 25±3°C. Allow at least 20 min-
utes to warm up. The DIP switch is located at the TC input.
Perform step 1 as stated above, then press the scroll key until the
display shows
.
Press the up and down keys until a value
of 0.00 is obtained. Press the scroll key at least 3 seconds. The
display will blink for a moment until a new value is obtained. If
the display didn’t blink or if the obtained value is equal to -5.00
or 40.00, then calibration failed.
• Perform step 11 to calibrate the gain of cold junction compen-
sation if required. If a test chamber for calibration is not avail-
able, perform step 11N to use a nominal value for the cold junc-
tion gain.
Step 11
Set up the equipment the same as in step 10. The unit under cal-
ibration is powered in a still-air room with a temperature of
50±3°C. Allow at least 20 minutes to warm up. The calibrator
source is set at 0.00°C with internal compensation mode.
Perform step 1 as stated above, then press the scroll key until
the display shows
.
Press the up and down keys until a
value of 0.0 is obtained. Press the scroll key for at least 3 sec-
onds. The display will blink for a moment until a new value is
obtained. If the display didn’t blink or if the obtained value is
equal to -199.9 or 199.9, then calibration failed. This setup is
performed in a high temperature chamber, therefore it is recom-
mended to use a computer to perform the procedures.
Step 11N
Perform step 1 as stated above, then press the scroll key until the
display shows . Press the up and down keys until a value
of 0.1 is obtained. Press the scroll key for at least 3 seconds. The
display will blink for a moment until the new value 0.0 is
obtained. Otherwise, calibration failed.
Caution: It is not recommended to use step 11N, since the
cold junction gain is not able to achieve the rated accuracy using
this step.
Final step
Step 12
Set the DIP switch to your desired position (refer to section 1-3).
Automatic calibration procedures
The programming port (see section 2-18) of the TEC-9300 can
be used for automatic calibration.
The equipment required for automatic calibration is available
upon request.
Manual calibration procedures, continued…