
Host High Speed Electrical Test Procedure Revision 1.4
7
on this computer is Windows 2000 or XP Professional. Please refer to the High-speed Electrical
Test Setup Instruction for steps to configure this computer.
Note: The Tektronix USB compliance test software operates internally to the oscilloscope and
automatically configures the oscilloscope settings so that external GPIB control of the
oscilloscope is not required. One can consolidate the High-speed Signal Quality Analysis
Computer and the High-speed USB Test Bed Computer into one single PC.
3.1 Equipment Setup
3.1.1 Test Equipment Setup Diskettes
Some tests described in this procedure are not automated by the Tektronix TDSUSB2
Compliance Test Software application. For those tests, the application includes setup files for the
test equipment used in this procedure. The setup files simplify equipment setup. Copy the HS
Scope Setup files to floppy disks. Insert the Tektronix TDS7000/6000, DPO7000/6000 setup
floppy diskette into the Tektronix TDS7254/7404, DPO7254/7354/7404, or TDS/DPO6000
oscilloscopes.
3.1.2 Digital Sampling Oscilloscope
Before turning on the oscilloscope, attach a P6330 or P6248 differential probe to Channel 1 and
Channel 4. TPA-BNC adaptors are required for the DPO7000 series. Attach two P6245 FET
probes, one to Channel 2 and one to Channel 3. The probe assignment will be used through out
the entire test procedure. Turn on the oscilloscope to allow for 20 minutes of warm up time prior
to use. Perform the signal path compensation procedure built into the Utility menu of the
Tektronix oscilloscope if the ambient temperature has changed more than 5 degrees. The
compensation should be performed with the probes disconnected from the oscilloscope.
A higher resolution view of the USB signaling is possible by turning on the scope’s Hi
Resolution mode. On the scope’s menu bar, select Horiz/acq > Horizontal/Acquisition Setup >
Acquisition > Hi Res.
The two single-end FET probes must be calibrated to minimize gain and offset errors. The offset
errors of the diff probes will be cancelled later as a part of the test procedure process. The offset
of the differential probe will be adjusted by the step identified in the test procedure.
For P6247/P6248 differential probes, the following setting will be used through out the entire test
procedure:
ο DC Reject <OFF> (P6247 only)
ο BW <Full> (P6247 only)
ο Attenuation <÷1>
The use of a Tip Saver accessory on the P6247 or P6248 probe is highly recommended to prolong
probe tip life. Please note that the Tip Saver will wear out after repeated use and start degrading
signal quality measurements. Replace the Tip Saver when measurements made without it provide
better results.
Note: In certain test situation, there may not be a ground connection between the DSO and the
device under test. This may lead to the signal seen by the differential probe to be modulated up
Comentários a estes Manuais