Commands Listed in Alphabetical Order
VOLTS, VOLTS SQ
UARED, SEC, HERTZ, PERCENT, DIVS, SAMPLES,
OHMS, AMPS, WATTS, MINUTES, DEGREES, UNKNOWN, AMPS
SQUARED, HOURS, DAYS, DB, BYTES, INVERSE HERTZ, IRE, V OVER V,
V OVER A, VOLTS WATTS, V OVER W, V OLTS DB, V OVER DB, A OVER
V, A OVER A, AMPS WATTS, A OVER W, AMPS DB, A OVER DB, WATTS
VOLTS, W OVER V, WATTS AMPS, W OVER A, WATTS SQUARED, W
OVER W, WATT
S DB, W OVER DB, DB VOLTS, DB OVER V, DB AMPS, DB
OVER A, DB WATTS, DB OVER W, DB SQUARED, DB OVER DB, VOLTS
SEC, AMPS SEC, WATTS SEC, V OVER S, A OVER S, W OVER S
Group
Measurement
Syntax
MEASUrement:IMMed:UNIts?
Examples
MEASUREMENT:IMMED:UNITS? might return
:MEASUREMENT:IMMED:UNIts "s"
indicating that units for the immediate measurement are in seconds.
MEASUrement:IMMed:VALue? (Query Only)
R
eturns the value of the measurement specied by the MEASUrement:IMMed:
TYPe command. The measurement is immediately taken on the source(s)
specied by a MEASU rement:IMMed:SOUrc e1 command.
NOTE. A change to HORizontal:MAIn:SCALe or CH<x>:SCALe will not
necessarily have taken affect if immediately followed by this command.
Group
Measurement
Syntax
MEASUrement:IMMed:VALue?
Related Commands
MEASUrement:IMMed:TYPe, MEASUrement:IMMed:SOUrce1 , *ESR?,
ALLEv?
Examples
MEASUREMENT:IMMED:VALUE? might return : MEASUREMENT:IMMED:VALUE
9.9000E+37.
If the measurement has an error or warning associated with it,
then an item is added
to the error queue. The error can be checked for with the
*ESR? and ALLEv? commands.
2-194 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual